Beckman Coulter Allegra X-15R Refrigerated Centrifuge
Temperature Range -10°C to 40°C
Maximum Capacity 4 x 750ml
- Equipped with a brushless asynchronous motor, ensuring efficiency and reducing maintenance needs.
- Automatic rotor identification system ensures safety by limiting the speed to the maximum safe speed of the identified rotor.
- Program memory enables repeated run conditions, allowing for consistent and repeatable operations.
- Includes an imbalance detector that monitors the rotor during operation, triggering automatic shutdown if rotor loads are severely out of balance.
- Electromechanical Door Lock System prevents operator contact with spinning rotors and ensures the centrifuge runs only when the door is shut.
The Beckman Coulter Allegra X-15R Refrigerated Centrifuge is a benchtop refrigerated centrifuge that generates centrifugal forces required for a wide variety of applications. The Allegra X-15R is microprocessor-controlled, providing interactive operation. The instrument design features a brushless asynchronous motor, automatic rotor identification system, program memory that enables repeated run conditions, temperature control system from -10°C to 40°C, and a choice of acceleration and deceleration rates. User messages alert the operator to conditions that may need attention. The Beckman Coulter Allegra X-15R Refrigerated Centrifuge also features an easy to use control panel, which is mounted at an angle on the front of the centrifuge for easy visibility and access. A rotor identification system prevents the installed rotor from running above its maximum rated speed. During acceleration the microprocessor checks the magnetic rotor identification. Speed is limited to the maximum safe speed of the identified rotor. If the system identifies a rotor different than the one entered by the user, the system displays an error message and reduces the speed to the maximum allowable speed of the rotor. An imbalance detector monitors the rotor during the run, causing automatic shutdown if rotor loads are severely out of balance. The included SX4750 Swing Bucket Rotor rotor reaches a maximum speed of 4,750 RPM and a maximum RCF of 5,250 x g. (Please Note: This centrifuge requires 208V for operation). All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
Please confirm this rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Beckman Coulter Allegra X-15R Refrigerated Centrifuge includes:
- > SX4750 Swing Bucket Rotor
- > Set of 4 x SX4750 Round Buckets
- > Set of 4 x Bucket Caps
- > Set of 15ml Adapters (2 Total)
- > Set of 50ml Adapters (2 Total)
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Beckman Coulter Allegra X-15R Refrigerated Centrifuge Specifications
Maximum Speed
|
10,200 RPM (4,750 RPM with included rotor)
|
Maximum RCF
|
11,400 x g (5,250 x g with included rotor)
|
Maximum Capacity
|
4 x 750ml
|
Temperature Range
|
-10°C to 40°C in 1°C Increments
|
Timer
|
Up to 9 hours 59 minutes, or continuous run (hold)
|
Acceleration Rates
|
10
|
Deceleration Rates
|
11
|
Noise Level
|
< 68 dBa
|
Maximum Heat Dissipation
|
6800 Btu/hr (1.2 kW)
|
Dimensions (W x D x H)
|
30 x 24.5 x 13.5 in. (76.2 x 62.2 x 34.3 cm)
|
Weight
|
283 lb. (128 kg)
|
Electrical Requirements
|
208V, 60Hz, 12A
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|