Barnstead Lab-Line MaxQ 5000 Refrigerated Incubator Shaker
The Barnstead Lab-Line MaxQ 5000 Refrigerated Incubator Shaker combines advanced functionality and reliability, making it ideal for applications including solubility studies, cell cultures, and general mixing. It offers a speed range of 15 to 500 RPM with precise control accuracy of ± 1 RPM, and a temperature range from Ambient -15°C to 60°C with an accuracy of ± 0.1°C at 37°C. The MaxQ 5000 features a large, easy-to-view LED display for monitoring temperature, speed, and time, allowing speed adjustments in 1 RPM increments and featuring an RS-232 interface for remote monitoring. Essential for continuous operations, it includes a triple eccentric drive that ensures smooth, uniform agitation even at high speeds and with unbalanced loads. The unit's safety features include a flexible door gasket, an overtemperature safety signal, and a safety interlock that halts operation when the door is opened. Also featured is a foot pedal for hands-free door operation. This incubator shaker will be fully tested to perform at factory specifications before being shipped.
The Barnstead Lab-Line MaxQ 5000 Refrigerated Incubator Shaker includes:
- > 18 x 18 in. Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Barnstead Lab-Line MaxQ 5000 Refrigerated Incubator Shaker Specifications
Speed Range
|
15 to 500 RPM
|
Speed Control Accuracy
|
± 1 RPM
|
Temperature Range
|
Ambient -15°C to 60°C
|
Temperature Accuracy
|
± 0.1°C at 37°C
|
Timer
|
Continuous or times from 0.1 to 999 hours or 0.1 to 999 minutes
|
Orbit
|
1 in. (2.54 cm)
|
Drive Mechanism
|
Triple-eccentric orbital drive with 6 permanently lubricated ball bearings
|
Display
|
LED
|
Platform Size (L x W)
|
18 x 18 in. (46 x 46 cm)
|
Overall Dimensions (W x H x L)
|
41 x 41 x 32 in. (104.1 x 104.1 x 81.3 cm)
|
Weight
|
512 lb. (232.2 kg)
|
Electrical Requirements
|
120V, 50/60Hz
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|