DAD Detector
|
Specification
|
Detection Type
|
1024-element photodiode array
|
Light Source
|
Deuterium and tungsten lamps
|
Wavelength Range
|
190 - 640 nm
|
Short Term Noise (ASTM) Single and Multi-Wavelength
|
± 3 × 10
-6 AU at 230 nm/4 nm
|
Drift
|
< 0.5 × 10
-3 AU/hr at 230 nm
|
Linear Absorbance Range
|
> 2.0 AU (5%) at 265 nm
|
Wavelength Accuracy
|
± 1 nm
|
Wavelength Bunching
|
1 - 400 nm
|
Slit Width
|
4 nm
|
Diode Width
|
~ 0.5 nm
|
Flow Cell
|
Standard: 13 µL volume, 10 mm cell path length and 120 bar (1740 psi) pressure maximum
|
Binary Pump
|
Specification
|
Hydraulic System
|
Two dual piston in series pumps with proprietary servo-controlled variable stroke drive, power transmission by gears and ball screws, floating pistons
|
Setable Flow Range
|
0.001 – 5 ml/min, in 0.001 ml/min increments
|
Flow Range
|
0.5 - 5.0 mL/min
|
Flow Precision
|
≤ 0.07% RSD or ≤ 0.02 min SD, whatever is greater
|
Flow Accuracy
|
± 1% or 10 µl/min, what ever is greater
|
Pressure Operating Range
|
Operating range 0 to 600 bar (0 – 7800 psi) up to 5 ml/min
|
Pressure Pulsation
|
< 2 % amplitude (typically < 1.3 %), or < 0.3 MPa (3 bar), whatever is greater, at 1 mL/min isopropanol, at all pressures > 1 MPa (10 bar, 147 psi)
|
Compressibility Compensation
|
User-selectable, based on mobile phase compressibility
|
Recommended pH Range
|
1.0 - 12.5, solvents with pH < 2.3 should not contain acids which attack stainless stee
|
Gradient Formation
|
High-pressure binary mixing
|
Delay Volume
|
600 to 800 µl, dependent on back pressure
|
Composition Range
|
0 to 100% (1 to 99% or 5 µl/min per channel, whatever is greater is recommended)
|
Composition Precision
|
< 0.15% RSD or < 0.04 min SD whatever is greater
|
Composition Accuracy
|
± 0.35% absolute at 2 mL/min, at 10 MPa (100 bar)
|
Autosampler
|
Specification
|
Injection Range
|
0.1 - 100 µl in 0.1 µl increments
|
Precision
|
< 0.25 % RSD of peak areas from 5 to 100 µl
< 1 % RSD of peak areas from 1 to 5 µl
|
Pressure Range
|
Operating range 0 - 60 MPa (0 - 600 bar, 0 - 8850 psi)
|
Sample Viscosity Range
|
0.2 to 50 cp
|
Sample Capacity
|
100 × 2 mL vials in 1 tray
40 × 2 mL vials in ½ tray 15 × 6 mL vials in ½ tray (Agilent vials only)
|
Injection Cycle Time
|
50 s for draw speed 200 µL/min, ejection speed 200 µL/min, injection volume 5 µL
|
Carryover
|
Typically < 0.1 %, < 0.05 % with external needle cleaning
|
ALS Thermostat
|
Specification
|
Temperature Range
|
4°C to 40°C in 1°C increments
|
Temperature Accuracy at Ambient Temperatures < 25°C and Humidity < 50%
|
3°C to +8°C at a setpoint of 4°C
|
Temperature Accuracy at Ambient Temperatures > 30°C and Humidity > 60%
|
3°C to +9°C at a setpoint of 4°C
|
Column Compartment
|
Specification
|
Temperature Range
|
Ambient -10°C to 80°C at flow rates up to 5 mL/min
|
Temperature Stability
|
± 0.15 °C
|
Temperature Accuracy
|
± 0.8°C (± 0.5°C with calibration)
|
Column Capacity
|
Three 30 cm
|
Warm-Up/Cool-Down Time
|
5 minutes from ambient to 40°C
10 minutes from 40°C to 20°C
|
Dead Volume
|
3 µl left heat exchanger
6 µl right heat exchanger
|
Degasser
|
Specification
|
Number of Solvent Channels
|
4
|
Maximum Flow Rate
|
0 - 5 mL/min per channel
|
Internal Volume Per Channel
|
Typically 1 ml per channel
|
Materials in Contact with Solvent
|
PTFE, PEEK
|
pH Range
|
1 - 14
|