Agilent 6130B Single Quadrupole Mass Spectrometer w/ 1260 Infinity I LC System
Mass Range 10 to 3,000 m/z
- Features 30 msec ultrafast ion polarity switching for maximum data collection
- Supports multiple ionization sources including ESI, APCI, APPI, and multimode options
- Provides dynamic range exceeding 6 × 10⁶ for reliable quantitation
- Ensures long-term stability with < 0.1 amu mass drift over 12 hours
- High-performance detector with high gain electron multiplier horn
- Fully controlled via Agilent ChemStation Workstation Software for streamlined operation
- SIM sensitivity with signal-to-noise ratio > 250:1 for ESI positive mode
- Maximum SIM dwell time of 5 msec for fast, accurate quantitation
- Equipped with a reliable vacuum system featuring a turbo-molecular and mechanical pump
The Agilent 6130B Single Quadrupole Mass Spectrometer w/ 1260 Infinity I LC System delivers unmatched analytical performance and proven day-after-day reliability from routine QC to research applications. This 6130B model is specifically a flexible, high performance solution ideal for any quantitation application with 3000 Da mass range and 1 pg sensitivity. This system allows users to rapidly screen compounds and confirm molecular weights, purify target compounds in complex mixtures, quantitate target compounds, positively identify impurities and more. The Agilent 6130B Single Quadrupole Mass Spectrometer provides unmatched sensitivity, simplifying sample preparation and LC method development. An ultrafast scan speed allows users to take full advantage of UHPLC separations, as well as featuring ultrafast ion polarity switching to return maximum information from a single injection, even with narrow LC peaks. The Agilent 6130B Single Quadrupole Mass Spectrometer w/ 1260 Infinity I LC System can be fully controlled from the included Agilent ChemStation Workstation Software, which allows for user-friendly operation in data acquisition, analysis, instrument control, automation, and more.
This turnkey Agilent LC/MS system was recertified and calibrated by a factory trained engineer and is ready to ship on demand. This system comes with everything needed for easy installation and setup. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Agilent 6130B Single Quadrupole Mass Spectrometer w/ 1260 Infinity I LC System includes:
- > Agilent 6130B Single Quadrupole Mass Spec
- > Agilent MS40+ Rotary Vane Vacuum Pump
- > G1948B ESI Ionization Source
- > G1312B Binary Pump (G1311B Quaternary Pump available, please inquire)
- > G1329B Autosampler
- > G1330B ALS Thermostat
- > G1316A TCC
- > G1379B Degasser
- > G4212B DAD Detector
- > Windows 10 Professional 64-Bit Data Station with Flatscreen Monitor
- > ChemStation Version 4.3 Software Installed
- > All Power Cords
- > Can/Crossover Cables
- > Agilent Lan Interface Card Built-In
- > Solvent Tubing
- > Capillaries
- > Agilent 4 Peak Test w/ Lab Advisor Diagnostics Report & Mass Spec Tuning Report
- > Full 180 Day Parts and Labor Warranty
Below is a List of Diagnostic Tests Performed On This System
- Filter Test: The test moves the filter into the light path and measures the relative absorbance in the filter position.
- Dark Current Test: The test measures the dark current from the detector optic.
- Holmium Oxide Test: The test measures the Holmium spectrum from the built-in Holmium filter. The spectrum is evaluated for peaks at different wavelengths.
- Lamp Intensity Test: The test verifies functionality of the UV lamp at various wavelengths.
- Wavelength Intensity Test: The test confirms the accuracy of wavelengths measured.
- Baseline Noise Test: The test measures the amount of baseline noise drift identified by the detector.
- Pressure Test: The pressure test is a quick, built-in test designed to demonstrate the pressure tightness (up to 400 bar) of the system.
- Leak Test: The leak test is a built-in troubleshooting designed to demonstrate the leak-tightness of the pump.
- Thermostat Test: Evaluates the cooling and heating performance of the two Peltier elements.
- Injector Steps: Performs each movement of the sampling sequence under manual control. This is useful for confirming autosampler functionality.
- 4 Peak Test (Isocratic Standard): Standard test published by Agilent to confirm complete functionality of the 1200 HPLC System.
Agilent 6130B Single Quadrupole Mass Spectrometer Specifications
Maximum Scan Speed
|
10,400 Da/sec
|
SIM Sensitivity, ESI Positive
|
Signal-to-noise ratio > 250:1 (1 pg of reserpine injected on column,
quantifying the m/z 609 ion)
|
SIM Sensitivity, APCI Positive
|
Signal-to-noise ratio > 100:1 (1 pg of reserpine injected on column,
quantifying the m/z 609 ion)
|
Polarity Switching
|
30 msec (switching time from positive to negative polarity)
|
Mass Resolution (Autotune)
|
Unit resolution (full width at half maximum)
|
Mass Stability
|
< 0.1 amu or 100 ppm, whichever is larger (over 12 hours)
|
Mass Accuracy
|
± 0.13 Da
|
Maximum SIM Dwell Time
|
5 msec
|
Maximum SIM Transitions
|
100 (one time segment)
|
Dynamic Range
|
> 6 × 10
6
|
Mass Range
|
10–3,000 m/z
|
Single Point of Control
|
Single-point data system method capability with full control of Agilent 1200 Series LC Systems and 6100 Series Single Quadrupole LC/MS Systems
|
Time Programming
|
SIM ion selections, Scan functions, Fragmentor voltage, External CAN valves
|
Ionization Sources
|
Electrospray (ESI), Multimode source (simultaneous ESI and APCI), Atmospheric pressure chemical ionization (APCI), ChipCube for intergrated nanoLC, Atmospheric pressure photoionization (APPI)
|
Detector
|
High energy conversion dynode and high gain electron multiplier horn
|
Vacuum System
|
One turbo-molecular pump with one mechanical pump
|
Dimensions (W x H x D)
|
14 x 18 x 25 in. (35 x 45 x 63.5 cm)
|
Weight
|
137.6 lb. (63.4 kg.)
|
Electrical Requirements
|
200-230VAC, 60Hz or 200-240VAC, 50Hz
|
Click Here to View Specifications for the Agilent 1260 Infinity I LC System
DAD Detector
|
Specification
|
Detection Type
|
1024-element photodiode array
|
Light Source
|
Deuterium and tungsten lamps
|
Wavelength Range
|
190 - 640 nm
|
Short Term Noise (ASTM) Single and Multi-Wavelength
|
± 3 × 10
-6 AU at 230 nm/4 nm
|
Drift
|
< 0.5 × 10
-3 AU/hr at 230 nm
|
Linear Absorbance Range
|
> 2.0 AU (5%) at 265 nm
|
Wavelength Accuracy
|
± 1 nm
|
Wavelength Bunching
|
1 - 400 nm
|
Slit Width
|
4 nm
|
Diode Width
|
~ 0.5 nm
|
Flow Cell
|
Standard: 13 µL volume, 10 mm cell path length and 120 bar (1740 psi) pressure maximum
|
Binary Pump
|
Specification
|
Hydraulic System
|
Two dual piston in series pumps with proprietary servo-controlled variable stroke drive, power transmission by gears and ball screws, floating pistons
|
Setable Flow Range
|
0.001 – 5 ml/min, in 0.001 ml/min increments
|
Flow Range
|
0.5 - 5.0 mL/min
|
Flow Precision
|
≤ 0.07% RSD or ≤ 0.02 min SD, whatever is greater
|
Flow Accuracy
|
± 1% or 10 µl/min, what ever is greater
|
Pressure Operating Range
|
Operating range 0 to 600 bar (0 – 7800 psi) up to 5 ml/min
|
Pressure Pulsation
|
< 2 % amplitude (typically < 1.3 %), or < 0.3 MPa (3 bar), whatever is greater, at 1 mL/min isopropanol, at all pressures > 1 MPa (10 bar, 147 psi)
|
Compressibility Compensation
|
User-selectable, based on mobile phase compressibility
|
Recommended pH Range
|
1.0 - 12.5, solvents with pH < 2.3 should not contain acids which attack stainless stee
|
Gradient Formation
|
High-pressure binary mixing
|
Delay Volume
|
600 to 800 µl, dependent on back pressure
|
Composition Range
|
0 to 100% (1 to 99% or 5 µl/min per channel, whatever is greater is recommended)
|
Composition Precision
|
< 0.15% RSD or < 0.04 min SD whatever is greater
|
Composition Accuracy
|
± 0.35% absolute at 2 mL/min, at 10 MPa (100 bar)
|
Autosampler
|
Specification
|
Injection Range
|
0.1 - 100 µl in 0.1 µl increments
|
Precision
|
< 0.25 % RSD of peak areas from 5 to 100 µl
< 1 % RSD of peak areas from 1 to 5 µl
|
Pressure Range
|
Operating range 0 - 60 MPa (0 - 600 bar, 0 - 8850 psi)
|
Sample Viscosity Range
|
0.2 to 50 cp
|
Sample Capacity
|
100 × 2 mL vials in 1 tray
40 × 2 mL vials in ½ tray 15 × 6 mL vials in ½ tray (Agilent vials only)
|
Injection Cycle Time
|
50 s for draw speed 200 µL/min, ejection speed 200 µL/min, injection volume 5 µL
|
Carryover
|
Typically < 0.1 %, < 0.05 % with external needle cleaning
|
ALS Thermostat
|
Specification
|
Temperature Range
|
4°C to 40°C in 1°C increments
|
Temperature Accuracy at Ambient Temperatures < 25°C and Humidity < 50%
|
3°C to +8°C at a setpoint of 4°C
|
Temperature Accuracy at Ambient Temperatures > 30°C and Humidity > 60%
|
3°C to +9°C at a setpoint of 4°C
|
Column Compartment
|
Specification
|
Temperature Range
|
Ambient -10°C to 80°C at flow rates up to 5 mL/min
|
Temperature Stability
|
± 0.15 °C
|
Temperature Accuracy
|
± 0.8°C (± 0.5°C with calibration)
|
Column Capacity
|
Three 30 cm
|
Warm-Up/Cool-Down Time
|
5 minutes from ambient to 40°C
10 minutes from 40°C to 20°C
|
Dead Volume
|
3 µl left heat exchanger
6 µl right heat exchanger
|
Degasser
|
Specification
|
Number of Solvent Channels
|
4
|
Maximum Flow Rate
|
0 - 5 mL/min per channel
|
Internal Volume Per Channel
|
Typically 1 ml per channel
|
Materials in Contact with Solvent
|
PTFE, PEEK
|
pH Range
|
1 - 14
|
Computer Specifications
|
Model
|
Dell Optiplex 7010 Desktop PC
|
Processor
|
Intel Core i5-6500 Processor 3.2GHz
|
RAM
|
16GB
|
Hard Drive
|
250GB Internal SSD
|
Operating System
|
Windows 10 Professional, 64bit
|
|
|
|
|
|
|
|