Agilent 2100 Bioanalyzer G2939B
Number of Samples Up to 12 Samples/Chips
Sizing Range 25 to 7000 bp
- Microfluidics-based platform integrating sample handling, separation, detection, and data analysis
- Electrophoresis-based assay principle providing flexibility and sensitivity down to pg/μL concentration
- Supports DNA fragment analysis, protein applications, and sampling QC in gene expression
- Processes up to 12 samples per chip with minimal sample consumption of 1 µL
- Analysis run time of 30 to 45 minutes for quick results
- Ambient operating temperature range of 5° to 40°C (41° to 104°F)
- Compact design with dimensions of 162 x 412 x 290 mm (6.4 x 16.2 x 11.4 in) and weight of 10 kg (22 lbs)
- Compatible with Agilent 2100 expert software for real-time data display and analysis
- Meets IEC, CSA, and UL safety standards for reliability and compliance
- Powered by 100-240VAC, 50-60Hz with low power consumption of 30W
The Agilent 2100 Bioanalyzer G2939B is a microfluidics-based platform that integrates sample handling, separation, detection, and data analysis, enabling prompt and accurate sizing, quantification, and quality control. The bioanalyzer operates on one assay principle – electrophoresis – offering maximum flexibility and sensitivity down to pg/μL concentration. The 2100 is an extremely easy to use analyzer; simply add the sample, start the chip run, and watch real-time data display. Agilent’s 2100 expert software provides qualitative and quantitative digital data for up to 12 samples in 30 minutes with minimal sample consumption.
Specific uses of the Agilent 2100 Bioanalyzer include DNA fragment analysis, protein applications, and sampling QC in gene expression. Agilent 2100’s applications consist of quality control, evaluation of integrity, concentration, and purity, and fluorescent analysis of intact cells of limited quantity. Please Note: The Chip Priming Station is NOT included standard and is required for operation. If you do not already have a Chip Priming Station, please select this option from the dropdown menu above.
The Agilent 2100 Bioanalyzer is fully tested to perform at factory specifications before being shipped.
The Agilent 2100 Bioanalyzer G2939B includes:
- > Windows 11 Data Station with Agilent 2100 B.02.12 SR1 Expert Software
- > Interface Cable
- > IKA Chip Shaker (Chip Priming Station & Test Chip Kit Upgrade Available)
- > Agilent 2100 Test Chip Calibration Report
- > All Power Cords
- > Full 180 Day Parts and Labor Warranty
Agilent 2100 Bioanalyzer G2939B Specifications
Analysis Run Time
|
30 to 45 min
|
Number of Samples
|
Up to 12 samples/chips
|
Sample Volume
|
1 µl
|
Sizing Range
|
25 to 7000 bp
|
Ambient Operating Temperature
|
5° to 40°C (41° to 104°F)
|
Safety Standards: IEC, CSA, UL
|
Installation Category II, Pollution Degree 2
|
Power Consumption
|
30W / 60 VA
|
Dimensions (W x D x H)
|
162 x 412 x 290 mm (6.4 x 16.2 x 11.4 in)
|
Electrical Requirements
|
100-240VAC, 50-60Hz
|
Computer Specifications
|
Processor
|
Intel Core I5-8350U CPU
|
RAM
|
16.0 GB
|
Hard Drive
|
512GB Internal SSD
|
Operating System
|
Windows 11 Professional, 64-bit
|
- Verify power supply functionality to ensure stable operation of all components
- Confirm precise seating and alignment of the electrode cartridge
- Detach and perform ultrasonic cleaning on electrode pin assembly
- Validate communication between analyzer and software for seamless integration
- Ensure Agilent lid opens and closes smoothly without abrasion against the casing
- Inspect internal wiring for any shorts, tears, or compromised insulation
- Check autofocus motors for full movement range and operational accuracy
- Execute comprehensive autofocus system diagnostic analysis
- Perform detailed diagnostics for electrode and diode functionality
- Complete remaining 2100 Expert diagnostic tests to verify optimal functionality
- Inspect power cords and interface cables for wear, tears, or exposed wiring
- Verify software is fully operational and compatible with the updated computer system
Technician Inspection and Recertification Checklist: Bioanalyzers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Verify power supply functionality
|
Completed
|
|
Confirm electrode cartridge seating and alignment
|
Completed
|
|
Clean electrode pin assembly using ultrasonic cleaning
|
Completed
|
|
Validate communication with software
|
Completed
|
|
Confirm lid opens and closes without abrasion
|
Completed
|
|
Inspect internal wiring for damage or shorts
|
Completed
|
|
Check autofocus motor functionality
|
Completed
|
|
Run autofocus diagnostic analysis
|
Completed
|
|
Perform electrode and diode diagnostics
|
Completed
|
|
Complete all 2100 Expert diagnostic tests
|
Completed
|
|
Inspect power and interface cables for damage
|
Completed
|
|
Confirm software and computer compatibility
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|