Home > Chromatography - HPLC >

Agilent 1260 Infinity II LC System w/ Vialsampler & DAD
Agilent 1260 Infinity II LC System w/ Vialsampler & DAD
 
 



Our Price: $54,995.00
Sale Price: $47,395.00
Money Saved: $7,600.00

Price with Selected Options: $47,395.00

Condition: Fully Recertified and Tested to Perform at Manufacturer Specifications
Warranty: 180-Day Full Parts and Labor (Click for Info)
Typically Ships Within: 3 to 5 Business Days
Included Documentation Agilent 4 Peak Test w/ Lab Advisor Diagnostics Report

Request a Formal Quote


Quantity in Stock:1

Product Code: AG-1260LCVSDADII


Select Included Option*:

Please Select an Option Below

Select Additional Detector*:


Qty:
Description
Recertification Protocol
Technical Specification Sheet
 

Agilent 1260 Infinity II LC System w/ Vialsampler & DAD



Wavelength Range (DAD)
190 to 950 nm

Flow Rate Range
0.001 to 5.0 ml/min

Pressure Range
Up to 60 MPa (600 bar, 8700 psi) up to 5 ml/min

Features
  • Provides liquid chromatograph offering ultra fast and high resolution separation capability for consistently accurate results
  • Delivers several ease-of-use features, making the 1260 Infinity II a standard system for routine analysis
  • Binary Pump ensures reproducible gradients and high-performance, ideal for routine applications with UV or MS detection (Quaternary Pump available, please inquire)
  • Integrated degasser in the Binary Pump with a volume of 0.45ml per channel helps maintain system stability and performance
  • Diode Array Detector (DAD) enhances light transmission to near 100% efficiency, reducing baseline drift and improving peak integration
  • Minimal carryover with the Vialsampler, ensuring high reliability for routine pharmaceutical, environmental, and food analyses
  • Fully controllable via Agilent OpenLab CDS Software for seamless operation in data acquisition, analysis, and instrument control
  • Allows for user configuration of standard or low delay volumes, tailoring the system to specific analytical needs
  • Equipped with spectral tools for comprehensive data analysis, including spectral libraries and peak purity functions

The Agilent 1260 Infinity II LC System w/ Vialsampler & DAD provides a liquid chromatograph offering ultra fast and high resolution separation capability for consistently accurate results. The 1260 Infinity II LC System delivers several ease-of-use features, making it a standard system for routine analysis. With an operating pressure of up to 600 bar and flow range from 0.001 to 5.0 ml/min, the 1260 Infinity II Binary Pump is suited for LC applications using conventional or superficially porous particle columns, as well as STM column technology. Gradient formation is based on a high-pressure mixing principle. Standard or low delay volumes can be easily configured by the user. The Agilent 1260 Infinity II Binary Pump offers reproducible gradients and high-performance, providing high-throughput and fast separations. The pump is suitable for routine applications, with UV or MS detection, where high speed and resolution with uncompromised data quality are required. A two-channel integrated degasser is located inside of the pump, which allows for a volume of 0.45ml per channel.

The Agilent 1260 Infinity II Vialsampler is an autosampler designed for the reliability and ease-of-use needed for routine pharmaceutical tasks and quality control, as well as for environmental and food analyses. It houses an integrated column oven for two LC columns up to 30 mm in length. The integrated column oven paired with a sample cooler allows for temperature control from 4°C to 80°C, all within one module. Two classic trays are included with the Vialsampler, which allows for a capacity of up to 100 x 2 mL vials. The Vialsampler allows for an injection range of 0.1 to 900 µl up to a pressure of 40 MPa.

The Agilent 1290 Infinity II Diode Array Detector (DAD) is based on the Agilent Max-Light cartridge cell with optofluidic waveguides that improve light transmission to near 100% efficiency without sacrificing resolution caused by cell dispersions effects. With typical detector noise levels of ± 0.6 x 10-6 AU the 60 mm flow cell gives up to 10 times higher sensitivity than detectors with conventional flow cells. Any compromising refractive index and thermal effects are almost eliminated entirely, resulting in significantly less baseline drift for more reliable and precise peak integration. For fast separations, this detector has multiple wavelength and full spectral detection at sampling rates up to 240 Hz. The long-life deuterium lamp allows for the highest intensity and lowest detection limits over a wavelength range of 190 to 950 nm.

The Agilent 1260 Infinity II LC System w/ Vialsampler & DAD can be fully controlled from the Agilent OpenLab CDS Software, which allows for user-friendly operation in data acquisition, analysis, instrument control, automation, and more.

This Agilent system is in excellent condition and will be fully tested to perform at factory specifications before being shipped.

INCLUDED:
The Agilent 1260 Infinity II LC System w/ Vialsampler & DAD includes:
  • > G7112B Binary Pump (With Integrated Degasser) [G7111B Quaternary Pump Upgrade Available, Please Inquire]
  • > G7115A DAD Detector
  • > G7129A Vialsampler (With Integrated Column Oven)
  • > Windows 11 Professional 64-Bit Data Station with Flatscreen Monitor
  • > OpenLab CDS Software Installed
  • > Can/Crossover Cables
  • > Agilent Lan Interface Card Built-In
  • > Solvent Tubing
  • > 4 x Bottles, Caps and Solvent Filters
  • > New Stainless Steel Capillaries
  • > 1/4" Wrench & Mobile Phase Priming Syringe
  • > Waste Tubing
  • > All Power Cords
  • > Agilent 4 Peak Test w/ Lab Advisor Diagnostics Report
  • > Full 180 Day Parts and Labor Warranty


Below is a List of Diagnostic Tests Performed On This System
G7115A Diode Array Detector Diagnostic Tests:
  • Filter Test: The test moves the filter into the light path and measures the relative absorbance in the filter position.
  • Dark Current Test: The test measures the dark current from the detector optic.
  • Holmium Oxide Test: The test measures the Holmium spectrum from the built-in Holmium filter. The spectrum is evaluated for peaks at different wavelengths.
  • Lamp Intensity Test: The test verifies functionality of the UV lamp at various wavelengths.
  • Wavelength Intensity Test: The test confirms the accuracy of wavelengths measured.
  • Baseline Noise Test: The test measures the amount of baseline noise drift identified by the detector.
G7112B Binary Pump Diagnostic Tests:
  • Pressure Test: The pressure test is a quick, built-in test designed to demonstrate the pressure tightness (up to 400 bar) of the system.
  • Leak Test: The leak test is a built-in troubleshooting designed to demonstrate the leak-tightness of the pump.
Integrated Column Oven Diagnostic Tests:
  • Thermostat Test: Evaluates the cooling and heating performance of the two Peltier elements.
G7129A Vialsampler Diagnostic Tests:
  • Injector Steps: Performs each movement of the sampling sequence under manual control. This is useful for confirming autosampler functionality.
System Tests:
  • 4 Peak Test (Isocratic Standard): Standard test published by Agilent to confirm complete functionality of the Agilent 1260 Infinity II LC System w/ Vialsampler & DAD.

Agilent 1260 Infinity II LC System w/ Vialsampler & DAD Specifications

DAD Detector Specification

Detection Type

1024-element photodiode array
Light Source Deuterium lamp
Number of Signals 8
Maximum Sampling Rate 240 Hz (both spectra and signals)
Wavelength Range 190 - 950 nm
Short Term Noise

With 10 mm Max-Light cartridge cell: < ± 3 x 10 -6 AU at 230/4 nm, slit width 4 nm, TC 2 s , ASTM

With 60 mm Max-Light cartridge cell (included): < ± 0.6 x 10 -6 AU at 230/4 nm, slit width 4 nm, TC 2 s , ASTM

Drift < 0.5 x 10 -3 AU/h at 230 nm
Linearity > 2.0 AU (5%) at 265 nm, typically 2.5 AU (5%)
Wavelength Accuracy ± 1 nm
Wavelength Bunching 2 - 400 nm in steps of 1 nm
Wavelength Precision < ± 0.1 nm
Slit Width 1, 2, 4, 8 nm
Diode Width ~0.5 nm
Flow Cell Max-Light cartridge cell (high sensitivity): 60 mm
Spectral Tools Data analysis software for spectra evaluation, including spectral libraries and peak purity functions

Click Here to View More Specifications for the Agilent 1260 Infinity II LC System w/ Vialsampler & DAD

Binary Pump Specification
Hydraulic System Two dual piston in series pumps with servo-controlled variable stroke drive, power transmission by gears and ball screws, floating pistons
Settable Flow Range 0.001 - 5.0 ml/min, in 0.001 ml/min increments
Recommended Flow Range

0.05 - 5.0 ml/min

Flow Precision ≤ 0.07% RSD or ≤ 0.02 min SD, whichever is greater
Flow Accuracy ± 1% or 10 µl/min, whichever is greater, pumping degassed H2O at 10 MPa (100 bar)
Pressure Operating Range

Up to 60 MPa (600 bar, 8700 psi) up to 5 ml/min

Pressure Pulsation < 2% amplitude (typically < 1.3%) or < 0.3 MPa (3 bar, 44 psi), whichever is greater
Compressibility Compensation Pre-defined, based on mobile phase compressibility
Recommended pH Range 1.0 - 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel
Gradient Formation High-pressure binary mixing
Delay Volume 600 - 900 µl, dependent on back pressure; measured with water at 1 ml/min (water/caffeine tracer)
Settable Composition Range 0 - 100%
Recommended Compostion Range 1 - 99% or 5µl/min per channel, whichever is greater
Composition Precision < 0.15% RSD or < 0.04 min SD, whatever is greater
Integrated Degassing Unit

Number of channels: 2

Internal volume per channel: 0.45 ml

Vialsampler Specification
Injection Range 0.1 - 900 µl in 0.1 µl increments with 900 µl up to 40 MPa
Injection Precision < 0.25% RSD of peak areas from 5 µl to 100 µl
Pressure Range 0 - 40 MPa (0 - 400 bar, 0 - 5801 psi)
Sample Viscosity Range 0.2 - 5 cp
Sample Capacity 100 x 2 mL vial (two classic trays included)
Injection Cycle Time

18 s for draw speed 200 µl/min

Ejection speed: 200 µl/min

Injection volume: 1 µl

Carry Over 1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in 300 µl microvial.
Integrated Column Oven Specification
Temperature Range Ambient +4°C to 80°C
Temperature Accuracy 2°C - 6°C at a setpoint of 4°C
Column Capacity 2 columns of up to 30 mm length