Home > Chromatography - HPLC >

New!  Agilent 1260 Infinity II HPLC G7114A VWD Detector
Agilent 1260 Infinity II HPLC G7114A VWD Detector
 
 

Stock Notification!

The Agilent 1200 HPLC G1315B DAD Detector you were interested in is currently unavailable, but we have found a great alternative at an excellent price!

Our Price: $8,495.00

Condition: Fully Recertified and Tested to Perform at Manufacturer Specifications
Warranty: 180-Day Full Parts and Labor (Click for Info)
Typically Ships Within: 1 to 3 Business Days
Included Documentation: Lab Advisor Calibration Report

Request a Formal Quote


Quantity in Stock:1

Product Code: AG-VWD1260II
Qty:

Description
Recertification Protocol
Technical Specification Sheet
 

Agilent 1260 Infinity II HPLC G7114A VWD Detector


The Agilent 1260 Infinity II HPLC G7114A Variable Wavelength Detector (VWD) delivers exceptional sensitivity and speed, making it an industry leader in its class. With time-programmable wavelength switching and dual-wavelength mode, this detector enhances sensitivity, selectivity, and data acquisition. Advanced hardware improvements ensure low detector noise (< ±2.5 AU) and minimal baseline drift (<1 x 10^-4 AU/h), enabling precise quantification of trace-level components. Featuring electronic temperature control (ETC) for maximum baseline stability under varying environmental conditions and a 120 Hz data acquisition rate for fast analysis, the detector supports high productivity and reliable quantification. Additional features include a wide linear range for analyzing primary compounds and impurities, RFID-enabled flow cells and lamps for enhanced traceability, and automated wavelength verification via a built-in holmium oxide filter. Maintenance is streamlined with front access to lamps and flow cells, extensive diagnostics, and early maintenance feedback (EMF) for tracking lamp usage, ensuring maximum uptime and efficiency.

All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.

INCLUDED:
The Agilent 1260 Infinity II HPLC G7114A VWD Detector includes:
  • > Firmware Version 07.02 (Can be changed upon request)
  • > Flow Cell
  • > Deuterium Lamp
  • > Lab Advisor Calibration Report
  • > Lan Cable / Power Cord
  • > Operator's Guide
  • > Full 180 Day Parts and Labor Warranty

Agilent 1260 Infinity II HPLC G7114A VWD Detector Specifications

Detection Type

Double-beam photometer
Light Source Deuterium lamp
Wavelength Range 190 - 600 nm
Noise

< ± 0.25 x 10^-5 AU, at 230 nm (single wavelength detection)

< ± 0.80 x 10^-5 AU, at 230 nm and 254 nm (dual wavelength detection)

Drift < 1 x 10^-4 AU/h at 230 nm
Linearity > 2.5 AU upper limit
Wavelength Accuracy ± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter
Wavelength Precision < ± 0.1 nm
Slit Width 6.5nm typical over whole wavelength range
Time Programmable Wavelength, polarity, peak width, lamp on/off
Flow Cell Standard: 14 µl volume, 10 mm cell path length and 40 bar (588 psi) pressure maximum
Spectral Tools Stop-flow wavelength scan
Dimensions (H x W x D) 5.5 x 15.6 x 17 in. (14 x 39.6 x 43.6 cm)
Weight 24.3 lb. (11 kg)
Electrical Requirements 100-240V, 50/60Hz, 80VA
Downloads


Share your knowledge of this product. Be the first to write a review »

Browse for more products in the same category as this item:

Chromatography - HPLC
Brand
Brand > Agilent
Chromatography Component
Chromatography Component > Detectors