Agilent 1260 Infinity II HPLC G7112B Binary Pump
The Agilent 1260 Infinity II HPLC G7112B Binary Pump is designed for high-performance liquid chromatography applications, delivering precise and reliable solvent delivery. With an operating pressure of up to 600 bar and a flow range of 0.001 to 5.0 mL/min, it supports conventional, superficially porous particle, and STM column technologies. The pump employs a two-channel, dual-piston in-series design for robust gradient formation via high-pressure mixing, ensuring accurate and reproducible gradients. Equipped with a two-channel integrated degasser (0.45 mL per channel), it offers exceptional flow stability, particularly at low flow rates, enhancing detector sensitivity. Key features include user-configurable standard or low delay volumes, a bypassable pulse damper and mixer for low flowrate applications, and a pressure sensor for continuous monitoring. Ideal for high-throughput and fast-gradient methods, the pump delivers unmatched speed and resolution for routine applications with UV or MS detection, ensuring uncompromised data quality. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Agilent 1260 Infinity II HPLC G7112B Binary Pump includes:
- > Firmware version 07.02 (Can be changed upon request)
- > Front Cover
- > Leak and Pressure Diagnostic Report
- > Lan Cable / Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Agilent 1260 Infinity II HPLC G7112B Binary Pump Specifications
Hydraulic System
|
Two dual piston in series pumps with servo-controlled variable stroke drive, power transmission by gears and ball screws, floating pistons
|
Settable Flow Range
|
0.001 - 5.0 ml/min, in 0.001 ml/min increments
|
Recommended Flow Range
|
0.05 - 5.0 ml/min
|
Flow Precision
|
≤ 0.07% RSD or ≤ 0.02 min SD, whichever is greater
|
Flow Accuracy
|
± 1% or 10 µl/min, whichever is greater, pumping degassed H2O at 10 MPa (100 bar)
|
Pressure Operating Range
|
Up to 60 MPa (600 bar, 8700 psi) up to 5 ml/min
|
Pressure Pulsation
|
< 2% amplitude (typically < 1.3%) or < 0.3 MPa (3 bar, 44 psi), whichever is greater
|
Compressibility Compensation
|
Pre-defined, based on mobile phase compressibility
|
Recommended pH Range
|
1.0 - 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel
|
Gradient Formation
|
High-pressure binary mixing
|
Delay Volume
|
600 - 900 µl, dependent on back pressure; measured with water at 1 ml/min (water/caffeine tracer)
|
Settable Composition Range
|
0 - 100%
|
Recommended Compostion Range
|
1 - 99% or 5µl/min per channel, whichever is greater
|
Composition Precision
|
< 0.15% RSD or < 0.04 min SD, whatever is greater
|
Integrated Degassing Unit
|
Number of channels: 2
Internal volume per channel: 0.45 ml
|
Dimensions (H x W x D)
|
7.1 x 15.6 x 17.2 in. (18 x 39.6 x 43.6 cm)
|
Weight
|
38.8 lb. (17.6 kg)
|
Electrical Requirements
|
100-240V, 50/60Hz, 90 VA
|
|
|
|
|
|
|
|