Delay Volume (in ul)
|
Fraction collector inlet to diverter valve: ~50 (typical, depends on length of the tubing)
Diverter valve: ~15
Diverter valve to needle: ~10
Needle: ~4
|
Maximum System Flow
|
10 mL/min
|
Maximum Collection Volume
|
~20 mL with 30 x 48 mm (OD x L) tube
~30 mL with 30 x 75 mm (OD x L) tube
|
Maximum Capacity
|
3 fraction collectors in parallel plus one recovery fraction collector
|
Cooling
|
Optional (with additional G1330B), performance depending on ambient conditions and the volume of collected fractions
|
Trigger Modes
|
Time slices
Peak (threshold, up- / downslope)
Timetable (combination of time intervals and peak)
Manual trigger (supported only with Agilent Instant Pilot G4208A)
|
Trigger Sources
|
G7115A, G7165A, G7114A, G6120BA, G6130BA, G7121A, G4260B, G7162A
|
Operating Modes
|
Discrete fractions: default mode for all vessels. The flow is diverted to waste, while moving from one vessel position to the next vessel position
Continuous flow: optional, available only when using deep well plates. It is possible to move from one well plate position to the next one without diverting the flow into the well plate to waste
Needle into location: Needle pushes into the vessel as deep as specified, for the use with capped vials and test tubes and well plates with closing mats
Droplet setup mode: enables the fraction collector to collect small fractions without bubbles. The tip of the fraction collector needle initially moves down to the bottom of the well. Then it slowly moves upwards while the fraction is collected.
|
Maximum Time to Move Between Neighboring Vessels
|
Movement in x-direction: < 0.15 s
Movement in y-direction: < 0.3 s
|
Diverter Valve
|
3/2 valve, with switching time < 100 ms
|
Maximum Pressure
|
6 bar at the diverter valve during switching
|
Plates/Trays
|
4 x well-plates full tray (MTP), only one type of well-plates can be used at a time in one tray
|
Fraction Containers
|
30 x 48 mm (OD x L) tubes, ~20 mL / tube
25 x 48 mm (OD x L) tubes, ~15 mL / tube
16 x 48 mm (OD x L) tubes, ~11 mL / tube
12 x 48 mm (OD x L) tubes, ~8 mL / tube
|
Maximum Tube Height
|
48 mm with long needle assembly G1367-87200
75 mm with short needle assembly G1364-87202
|
Dimensions (H x W x D)
|
8 x 13.5 x 17 in. (20 x 34.5 x 44 cm)
|
Weight
|
29.8 lb. (13.5 kg)
|
Electrical Requirements
|
100-240V, 50/60Hz
|