|
Agilent 1260 Infinity I HPLC System
Wavelength Range (DAD) 190 to 640 nm
Flow Rate Range 0.001 to 10 mL/min
Pressure Range 0 to 600 bar (0 to 60 MPa) at 5 mL/min
- Offers exceptional flexibility for solvent selection and automation in HPLC method development and research
- Suited for multi-method, high-throughput workflows with easy access to four solvents for isocratic or gradient analysis
- Features a quaternary pump with a pressure capacity up to 600 bar for efficient solvent delivery
- Includes the 1260 Infinity Autosampler for low carryover and precise variable volume injection
- Equipped with a Diode-Array Detector (DAD) with a wavelength range of 190 - 640 nm for high optical performance
- Controlled by Agilent ChemStation Workstation Software for easy data acquisition, analysis, and instrument control
- Autosampler offers a broad injection range and sample capacity, accommodating various sample types and volumes
- Thermostatted column compartment and ALS thermostat provide precise temperature management for retention time reproducibility
The Agilent 1260 Infinity I HPLC System offers
the most flexibility for solvent selection and automation in HPLC method development, research and all HPLC applications requiring continuous access to a wide range of solvent choices. This system is ideally suited for multi-method, high-throughput workflows and offers convenient access to four solvents for isocratic or gradient analysis, for rapid method development and speeding up preparation of mobile phases and flushing the HPLC system.
The Agilent 1260 Infinity HPLC System's quaternary pump is based on a two-channel, dual-plunger in-series design which comprises all essential functions that a solvent delivery system has to fulfill. Metering of solvent and delivery to the high-pressure side are performed by one pump assembly which can generate pressure up to 600 bar. Degassing of the solvents is done in a built-in vacuum degasser integrated into the quaternary pump. The 1260 Infinity Autosampler is designed to offer the well-established flow-through design with variable volume injection and to achieve extremely low carryover, providing increased analysis speed with sensitivity, resolution, and precision. Paired with the autosampler is a thermostatted column compartment and ALS thermostat to serve as a temperature management system providing heating and cooling in order to meet extreme requirements of retention time reproducibility. The featured Diode-Array Detector (DAD) is designed for the highest optical performance, GLP compliance, and easy maintenance. The DAD provides greatly enhanced sensitivity and baseline stability using a cartridge cell system with optofluidic waveguides. This detector provides for a wavelength range of 190 - 640 nm, an 80 Hz data acquisition rate, and fixed 4 nm slit for rapid optimization of sensitivity, linearity, and spectral resolution.
The Agilent 1260 Infinity HPLC System can be fully controlled from the included Agilent ChemStation Workstation Software, which allows for user-friendly operation in data acquisition, analysis, instrument control, automation, and more.
This Agilent system is in excellent condition and will be fully tested to perform at factory
specifications before being shipped.
The Agilent 1260 Infinity I HPLC System includes:
- > G1311B Quaternary Pump (With Built-In Degasser)
- > G1329B Autosampler
- > G1330B ALS Thermostat
- > G1316C TCC
- > G1315C DAD Detector
- > Windows 11 Professional 64-Bit Data Station with Flatscreen Monitor
- > ChemStation Version 4.3 Software Installed
- > Can/Crossover Cables
- > Agilent Lan Interface Card Built-In
- > Solvent Tubing
- > 4 x Bottles, Caps and Solvent Filters
- > New Stainless Steel Capillaries
- > 1/4" Wrench & Mobile Phase Priming Syringe
- > Waste Tubing
- > All Power Cords
- > Agilent 4 Peak Test w/ Lab Advisor Diagnostics Report
- > Full 180 Day Parts and Labor Warranty
Below is a List of Diagnostic Tests Performed On This System
- Filter Test: The test moves the filter into the light path and measures the relative absorbance in the filter position.
- Dark Current Test: The test measures the dark current from the detector optic.
- Holmium Oxide Test: The test measures the Holmium spectrum from the built-in Holmium filter. The spectrum is evaluated for peaks at different wavelengths.
- Lamp Intensity Test: The test verifies functionality of the UV lamp at various wavelengths.
- Wavelength Intensity Test: The test confirms the accuracy of wavelengths measured.
- Baseline Noise Test: The test measures the amount of baseline noise drift identified by the detector.
- Pressure Test: The pressure test is a quick, built-in test designed to demonstrate the pressure tightness (up to 400 bar) of the system.
- Leak Test: The leak test is a built-in troubleshooting designed to demonstrate the leak-tightness of the pump.
- Thermostat Test: Evaluates the cooling and heating performance of the two Peltier elements.
- Injector Steps: Performs each movement of the sampling sequence under manual control. This is useful for confirming autosampler functionality.
- 4 Peak Test (Isocratic Standard): Standard test published by Agilent to confirm complete functionality of the 1260 Infinity HPLC System.
Agilent 1260 Infinity I HPLC System Specifications
DAD Detector
|
Specification
|
Detection Type
|
1024-element photodiode array
|
Light Source
|
Deuterium lamp
|
Wavelength Range
|
190 - 640 nm
|
Short Term Noise (ASTM) Single and Multi-Wavelength
|
< ± 3 × 10^-6 AU at 230 nm/4 nm
|
Drift
|
< 0.5 × 10^-3 AU/hr at 230 nm
|
Linear Absorbance Range
|
> 2.0 AU (5%) at 265 nm
|
Wavelength Accuracy
|
± 1 nm
|
Wavelength Bunching
|
2 - 400 nm
|
Slit Width
|
Fixed 4 nm slit
|
Diode Width
|
~ 0.5 nm
|
Signal Data Rate
|
80 Hz
|
Spectra Data Rate
|
80 Hz
|
Flow Cells
|
Max-Light Cartridge Cell (10 mm, V [σ], 1.0ul), 60 bar (870 psi) pressure maximum
Max-Light Cartridge Cell (60 mm, V [σ], 4.0ul), 60 bar (870 psi) pressure maximum
Max-Light Cartridge Test Cell
|
Control and Data Evaluation
|
Agilent ChemStation for LC
|
Analog Outputs
|
Recorder/integrator: 100 mV or 1 V, output range 0.001 - 2 AU, one output
|
Communications
|
Controller-area network (CAN), GPIB,
RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN
|
Safety and Maintenance
|
Extensive diagnostics, error detection, and display (through control module
and ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
|
GLP Features
|
Early maintenance feedback (EMF) for continuous tracking of instrument
usage in terms of lamp burn time with user-settable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter.
|
Housing
|
All materials recyclable
|
Click Here to View More Specifications for the Agilent 1260 Infinity HPLC System
Quaternary Pump
|
Specification
|
Setable Flow Range
|
0.001 - 10 mL/min, in 0.001 mL/min increments
|
Flow Range
|
0.05 - 10.0 mL/min
|
Flow Precision
|
< 0.07% RSD or < 0.02 min SD, whatever is greater based on retention time at constant ambient temperature
|
Flow Accuracy
|
± 1% or 10 µl/min
|
Pressure Operating Range
|
0 - 60 MPa (0 - 600 bar, 0 - 8,700 psi) up to 5 mL/min, 0 - 20 MPa (0 - 200 bar, 0 - 2950 psi) up to 10 mL/min
|
Pressure Pulsation
|
< 2% amplitude or < 3 bar at 1 mL/min isopropanol, at all pressures > 1 MPa (10 bar, 147 psi)
|
Composition Range
|
Settable range: 0 - 95% or 5 - 100%
|
Composition Precision
|
< 0.2 % RSD or < 0.04 min SD
|
Delay Volume
|
600 - 900 µl, dependent on back pressure
|
Autosampler
|
Specification
|
Injection Range
|
0.1 - 100 µl in 0.1 µl increments
|
Precision
|
< 0.25% from 5 - 40µl
< 0.5% from 2 - 5µl < 0.7% from 1 - 2µl < 1.5% from 0.5 - 1µl
|
Pressure Range
|
Up to 600 bar (8,700 psi)
|
Sample Viscosity Range
|
0.2 - 5 cp
|
Sample Capacity
|
2 x well plates (MTP) + 10 x 2 ml vials
108 x 2 ml vials in 2 x 54 vial plate plus 10 additional 2 ml vials 30 x 6 ml vials in 2 x 15 vial plate 100 Micro vial tray plus 10 additional 2 ml vials 54 Eppendorf tubes (0.5/1.5/2 ml) in 2 x 27 Eppendorf tube plates
|
Injection Cycle Time
|
Typically < 21 seconds using default conditions and injection volume of 5µl
|
Carryover
|
Typically < 0.004%
|
ALS Thermostat
|
Specification
|
Temperature Range
|
4°C to 40°C in 1°C increments
|
Temperature Accuracy at Ambient Temperatures < 25°C and Humidity < 50%
|
3°C to 8°C at a setpoint of 4°C
|
Temperature Accuracy at Ambient Temperatures > 30°C and Humidity > 60%
|
3°C to 9°C at a setpoint of 4°C
|
Column Compartment
|
Specification
|
Temperature Range
|
Ambient -10°C to 80°C at flow rates up to 5 mL/min
|
Temperature Stability
|
± 0.15°C
|
Temperature Accuracy
|
± 0.8°C
|
Warm-Up/Cool-Down Time
|
5 minutes from ambient to 40°C
10 minutes from 40°C to 20°C
|
Computer Specifications
|
Processor
|
Intel Core I5-8500 CPU
|
RAM
|
32.0 GB
|
Hard Drive
|
500GB Internal SSD
|
Operating System
|
Windows 11 Professional, 64-bit
|
- Verify uniformity of main and residential firmware across all components, updating as necessary to ensure compatibility
- Confirm successful communication between software and the HPLC system, validating data transfer integrity
- Check the degasser’s internal vacuum to confirm it reaches the specified vacuum level for optimal performance
- Replace inlet, outlet, purge valve, and pump seals in the system pump to maintain reliable operation
- Ensure system pump passes pressure and leak tests, confirming secure and efficient performance
- Replace all capillary tubing with new OEM stainless steel tubing for durability and precision
- Remove autosampler transport assembly, replace X and Y bearings, and thoroughly clean carrier rails to remove debris and contaminants
- Replace metering valve seal, autosampler needle, and needle seat for consistent sampling accuracy
- Test column compartment switching valve functionality and replace its seal if necessary
- Confirm column compartment passes thermostat test to maintain precise temperature control
- Remove and clean internal detector optics for clear and accurate signal detection
- Replace UV and Vis detector lamps to ensure optimal light source stability and performance
- Verify flow cell absorbance meets manufacturer’s specifications for accurate measurement
- Confirm system response to isocratic test sample, ensuring stable and reproducible results
Technician Inspection and Recertification Checklist: HPLC Systems
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Verify firmware uniformity and update as necessary
|
Completed
|
|
Confirm software communication with HPLC system
|
Completed
|
|
Check degasser vacuum level for optimal performance
|
Completed
|
|
Replace inlet, outlet, purge valve, and pump seals
|
Completed
|
|
Ensure pump passes pressure and leak tests
|
Completed
|
|
Replace all capillary tubing with OEM stainless steel tubing
|
Completed
|
|
Remove autosampler transport assembly, replace bearings, and clean rails
|
Completed
|
|
Replace metering valve seal, autosampler needle, and needle seat
|
Completed
|
|
Test column compartment switching valve and replace seal if needed
|
Completed
|
|
Confirm thermostat test in column compartment
|
Completed
|
|
Clean internal detector optics
|
Completed
|
|
Replace UV and Vis detector lamps
|
Completed
|
|
Verify flow cell absorbance meets manufacturer’s specifications
|
Completed
|
|
Confirm system response to isocratic test sample
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|