Delay Volume (in ul)
|
Fraction collector inlet to diverter valve: ~50 (typical, depends on length of the tubing)
Diverter valve: ~15
Diverter valve to needle: ~10
Needle: ~4
|
Maximum System Flow
|
10 mL/min (depending on viscosity and generated back pressure, max. 6 bar at the diverter valve). The analytical scale fraction collector can be modified for flow rates > 10 ml/min
|
Maximum Collection Volume
|
~20 mL with 48 mm tubes
~30 mL with 75 mm tubes
|
Maximum Capacity
|
3 fraction collectors in parallel plus one recovery fraction collector connected via 12-Position, 13-Port Selector valve (PN G1160A)
|
Cooling
|
Optional (with additional G1330B), performance depending on ambient conditions and the volume of collected fractions
|
Trigger Modes
|
Time slices
Peak (threshold, up- / downslope)
Timetable (combination of time intervals and peak)
Manual trigger (supported only with Agilent Instant Pilot G4208A)
|
Trigger Sources
|
Agilent 1260 Infinity DAD/MWD detectors (G1315C,D; G4212B/ G1365 C,D), the Agilent 1260 Infinity Fluorescence Detector (G1321A) and the Agilent G1946C/D, G1956A/B LC-MSD are fully supported
|
Operating Modes
|
Discrete fractions: default mode for all vessels. The flow is diverted to waste, while moving from one vessel position to the next vessel position
Continuous flow: optional, available only when using deep well plates. It is possible to move from one well plate position to the next one without diverting the flow into the well plate to waste
Needle into location: Needle pushes into the vessel as deep as specified, for the use with capped vials and test tubes and well plates with closing mats
Droplet setup mode: The tip of the fraction collector needle will initially move down to the bottom of the well. Then it will slowly move upwards while the fraction is collected. The droplet setup mode enables the fraction collector to collect small fractions without bubbles
|
Diverter Valve
|
3/2 valve, with switching time < 100 ms
|
Maximum Pressure
|
6 bar at the diverter valve during switching
|
Plates/Trays
|
4 x well-plates full tray (MTP)*
2 x well-plates std. tray + 10 funnels with external containers* (+ 1 half tray)
2 × well-plates std. tray (MTP) + 10 × 2 ml vials* (+ 1 half tray)
100 x 2 ml in std. tray (+ 1 half tray)*
3 x 40 x 2 ml in half tray*
3 x 40 funnels in half tray
3 x 15 x 6 ml in half tray*
Full tray with 40 test tubes (30 mm OD, max. height 48 mm, ~20 ml vol.)
Full tray with 60 test tubes (25 mm OD, max. height 48 mm
Full tray with 126 test tubes (16 mm OD, max. height 48 mm)
Full tray with 215 test tubes (12 mm OD, max. height 48 mm)
|
Maximum Tube Height
|
48 mm with long needle assembly G1367-87200
75 mm with short needle assembly G1364-87202
|
Delay Calibration Sensor
|
Single wavelength absorbance detector working at 654 nm, consisting of a LED and a photo diode
|
Diverter Valve
|
3/2 Diverter Valve with low internal volume (15 ul), switching time < 100 ms, maximum operating pressure 6 bar
|
Dimensions (H x W x D)
|
8 x 13.5 x 17 in. (20 x 34.5 x 44 cm)
|
Weight
|
29.8 lb. (13.5 kg)
|
Electrical Requirements
|
100-240V, 50/60Hz, 200VA
|