Agilent 1260 Infinity I HPLC G1329B Autosampler
Injection Range 0.1 – 100 µl
Injection Precision
0.25 % RSD of peak areas from 5 to 100 µl
Pressure Range
0 - 60 Mpa
- Transport mechanism uses an X-Z-Theta movement to optimize vial pick-up and return.
- Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display.
- Early maintenance feedback (EMF), electronic records of maintenance and errors.
- The autosampler continuous flow-through design ensures that sample loop, needle inside, seat capillary, and the mainpass of the injection valve is always in the flow line.
The Agilent 1260 Infinity I HPLC G1329B Autosampler is designed to offer the well-established flow-through design with variable volume injection and to achieve extremely low carryover, providing increased analysis speed with sensitivity, resolution, and precision. This autosampler features a sample capacity of up to 100 samples. The 1260 Infinity G1329B Autosampler is designed for use with other modules of the Agilent 1200 Infinity Series, 1200 Series, or 1100 Series LC, or with other LC systems if adequate remote control inputs and outputs are available. The autosampler is controlled by the Agilent Instant Pilot or by Agilent control software The G1329B's transport mechanism uses an X-Z-Theta movement to optimize vial pick-up and return. A gripper arm picks up the vials and positions them below the sampling unit. Movement is also monitored by optical sensors and optical encoders to ensure correct operation. The metering device is always flushed after injection to ensure minimum carry-over. The standard analytical head of the metering device provides injection volumes from 0.1 to 100 µl, and can be operated up to 600 bar. For applications that require temperature control, this module can be paired with the Agilent 1290 Infinity Thermostat (G1330B). All of our equipment is fully tested to perform at factory specifications before being shipped.
The Agilent 1260 Infinity I HPLC G1329B Autosampler includes:
- > Firmware Version 07.02 (Can be changed upon request)
- > Front Cover
- > 100 Vial Tray
- > Operator's Guide
- > Lab Advisor Documentation
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Agilent 1260 Infinity I HPLC G1329B Autosampler Specifications
Pressure
|
Operating range 0 - 60 MPa (0 - 600 bar, 0 - 8850 psi)
|
GLP Features
|
Early maintenance feedback (EMF), electronic records of maintenance
and errors
|
Communications
|
Controller-area network (CAN), RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
|
Safety Features
|
Leak detection and safe leak handling, low voltages in maintenance
areas, error detection and display
|
Injection Range
|
0.1 – 100 µl in 0.1 µl increments (recommended 1
µl increments)
Up to 1500
µl with multiple draw (hardware modification required)
|
Replicate Injections
|
1 – 99 from one vial
|
Precision
|
Typically < 0.25 % RSD of peak areas from 5 to 100 µl
Typically < 1 % RSD of peak areas from 1 to 5 µl
|
Minimum Sample Volume
|
1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample
in 300 µl microvial
|
Carryover
|
Typically < 0.1 %, < 0.05 % with external needle cleaning
|
Sample Viscosity Range
|
0.2 – 50 cp
|
Metering Devices
|
Metering pump in high pressure flow path
|
Sample Capacity
|
100 × 2 ml vials in 1 tray
40 × 2 ml vials in ½ tray
15 × 6 ml vials in ½ tray (Agilent vials only)
|
Injection Cycle Time
|
50 seconds for draw speed 200
µl/min, ejection speed 200 µl/min, injection volume 5 µl
|
Ambient Operating Temperature
|
0 to 55°C
|
Dimensions (H x W x D)
|
8 x 13.5 x 17 in (20 x 34.5 x 43.5 cm)
|
Weight
|
32 lb. (14.2 kg)
|
|
|
|
|
|
|
|