Hydraulic System
|
Dual piston in series pump with servo-controlled variable stroke drive, power transmission by gears and ball screws, floating pistons
|
Settable Flow Range
|
Set points 0.001 to 10 ml/min, in 0.001 ml/min increments
|
Flow Range
|
0.2 to 10.0 ml/min
|
Flow Precision
|
≤ 0.07% RSD, or ≤ 0.02 min SD whatever is greater, based on retention time at constant room temperature
|
Flow Accuracy
|
± 1% or 10 µl/min, whatever is greater, pumping degassed H2O at 10 MPa (100 bar)
|
Pressure Operating Range
|
Operating range up to 40 MPa (400 bar, 5880 psi) up to 5 ml/min
Operating range up to 20 MPa (200 bar, 2950 psi) up to 10 ml/min
|
Pressure Pulsation
|
< 2 % amplitude (typically < 1.0%) or < 0.3 MPa (3 bar, 44 psi), whatever is greater, at 1 ml/min isopropanol, at all pressures > 1 MPa (10 bar, 145 psi)
|
Compressibility Compensation
|
User selectable, based on mobile phase compressibility
|
Recommended pH Range
|
1.0 to 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel
|
Gradient Formation
|
Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve
|
Delay Volume
|
600 to 900 µl, dependent on back pressure; measured with water at 1 ml/min
|
Composition Range
|
0 to 95% or 5 to 100%, user selectable
|
Composition Precision
|
< 0.2 % RSD or < 0.04 min SD, whatever is greater, at 1 ml/min; based on retention time at constant room temperature
|
Integrated Degasser
|
Number of channels: 4
Internal volume per channel: 1.5 ml
|
Dimensions (H x W x D)
|
7 x 13.5 x 17 in. (18 x 34.5 x 43.5 cm)
|
Weight
|
32 lb. (14.5 kg.)
|
Electrical Requirements
|
100 - 240V, 50/60Hz, 180VA
|