Detection Type
|
Multi-signal fluorescence detector with rapid on-line scanning
capabilities and spectral data analysis
|
Performance Specifications
|
10 fg Anthracene, Ex=250 nm, Em=400 nm* RAMAN single wavelength (H2O)
> 500 with Ex=350 nm, Em=397 nm, dark value 450 nm, standard flow cell
time constant=4 seconds (8 seconds responsetime) RAMAN dual wavelength (H2O)
> 300 with Ex=350 nm, Em=397 nm, dark value 450 nm, standard flow cell
time constant=4 seconds (8 seconds responsetime)
|
Light source
|
Xenon Flash Lamp, normal mode 20 W, economy mode 5 W
|
Pulse frequency
|
296 Hz for single signal mode 74 Hz for spectral mode
|
Excitation Monochromator
|
Range: 200 nm - 700 nm and zero-order Bandwidth: 20 nm (fixed)
Monochromator:concave holographic grating, F/1.6, blaze: 300 nm
|
Emission Monochromator
|
Range: 280 nm - 900 nm and zero-order Bandwidth: 20 nm (fixed)
Monochromator:concave holographic grating, F/1.6, blaze: 400 nm
|
Reference System:
|
in-line excitation measurement
|
Timetable programing
|
up to 4 signal wavelengths, response time, PMT Gain, baseline
behavior (append, free, zero), spectral parameters
|
Spectrum acquisition:
|
Excitation or Emission spectra Scan speed: 28 ms per datapoint (e.g.
0.6 s/spectrum 200-400 nm, 10 nm step) Step size: 1-20 nm Spectra storage:
All
|
Wavelength characteristic
|
Repeatability+/- 0.2 nm Accuracy+/- 3 nm setting
|
Flow cells
|
Standard: 8 µl volume and 20 bar (2 MPa) pressure maximum, quartz
Optional: Fluorescence cuvette for offline spectroscopic measurements with 1
ml syringe, 8 µl volume, quartz
|
Control and data evaluation
|
Agilent ChemStation for LC, Agilent Instant Pilot G4208A or Agilent
Control Module G1323B with limited spectral data analysis and printing of
spectra
|
Communications
|
Controller-area network (CAN), GPIB, RS-232C, LAN, APG Remote: ready,
start, stop and shut-down signals
|