Agilent 1200 HPLC G1315D Diode Array Detector
Wavelength Range 190 to 950 nm
Wavelength Accuracy ± 1 nm
Drift 2 × 10-3 AU/hr at 254 nm
- Diode Array Detector with 1024-element photodiode array for precise detection
- Wide wavelength range from 190 to 950 nm for versatile applications
- Combined deuterium and tungsten lamps ensure high intensity and low detection limits
- Programmable slit width from 1 – 16 nm for tailored sensitivity and spectral clarity
- Baseline stability maintained through advanced temperature management
- Short-term noise as low as ±1 × 10⁻⁵ AU, offering high measurement accuracy
- Multiple flow cell options for varied applications, including high-pressure and nano-scale analysis
- Automated wavelength verification and GLP traceability with holmium oxide filter
- Analog outputs and multi-channel connectivity including CAN, GPIB, and RS-232C
- Early Maintenance Feedback (EMF) for tracking usage and alerts on lamp life
The Agilent 1200 G1315D Diode Array Detector verifies separation
quality with peak purity, multiple signals, and spectral libraries during analytical quantification, offering an extended wavelength range with the
highest sensitivity. The diode array detector allows for automated
wavelength verification for performance tracking. The Agilent 1200’s temperature
management system ensures optimum baseline stability, even under rough ambient
conditions. This detector features superb illumination using combined deuterium
and tungsten lamps for the highest intensity and lowest detection limit from
190 to 950 nm, 1024 diodes and 1-nm slit for highest spectral resolution. The
programmable slit helps you explore the spectral landscape for faster sample
characterization, ensuring higher sensitivity or lower baseline noise as necessary
on the 1100. Slit settings can be stored and reported from 1 – 16 nm with raw
data for GLP traceability.
The Agilent G1315D DAD is in excellent condition and will be fully tested to perform at factory
specifications before being shipped.
The Agilent 1200 HPLC G1315D DAD Detector includes:
- > Firmware Version 07.02 (Can be changed upon request)
- > Flow Cell
- > Tungsten and Deuterium Lamps
- > Lab Advisor Calibration Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Agilent 1200 HPLC G1315D Diode Array Detector Specifications
Type
|
Specification
|
Detection type
|
1024-element photodiode array
|
Light Source
|
Deuterium and tungsten lamps
|
Wavelength range
|
190 – 950 nm
|
Short term noise (ASTM) single and multi-Wavelength
|
± 1 × 10-5 AU at 254 and 750 nm
|
Drift
|
2 × 10-3 AU/hr at 254 nm
|
Linear absorbance
range
|
> 2 AU (upper limit)
|
Wavelength Accuracy
|
± 1 nm
|
Wavelength bunching
|
1 – 400 nm
|
Slit width
|
1, 2, 4, 8, 16 nm
|
Diode Width
|
< 1 nm
|
Flow Cells
|
Standard: 13 µl volume, 10 mm cell path length and 120 bar (1760 psi)
pressure maximum Semi-Micro: 5 µl volume, 6 mm cell path length and 120 bar
(1760 psi) pressure maximum High pressure: 1.7 µl volume, 6 mm cell path
length and 400 bar (5880 psi) pressure maximum 500 nano: 0.5 µl volume, 10 mm
cell path length and 50 bar (725 psi) pressure maximum
|
Control and data evaluation
|
Agilent ChemStation for LC
|
Analog outputs
|
Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, two
outputs
|
Communications
|
Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready,
start, stop and shut-down signals, LAN optional
|
Safety and maintainence
|
Extensive diagnostics, error detection and display (through control
module and ChemStation), leak detection, safe leak handling, leak output
signal for shutdown of pumping system. Low voltages in major maintenance
areas
|
GLP features
|
Early maintenance feedback (EMF) for continuous tracking of
instrument usage in terms of lamp burn time with user-settable limits and
feedback messages. Electronic records of maintenance and errors. Verification
of wavelength accuracy with built-in holmium oxide filter.
|
Housing
|
All materials recyclable
|
|
|
|
|
|
|
|