Agilent 1200 HPLC G1312B SL Binary Pump
Setable Flow Range Setpoints 0.001 – 5 ml/min, in 0.001 ml/min increments
Flow Range 0.1 – 5.0 ml/min
Pressure Range 0 - 400 bar (0 – 5880 psi) up to 5 ml/min
- Features a two-channel, dual-piston in-series design.
- Comprises all essential functions of a solvent delivery system.
- Metering of solvent and delivery to the high-pressure side are performed by two pump assemblies.
- Variable stroke volume allows pulse-free solvent delivery and efficient mixing.
- Active inlet valve synchronized to piston stroke eliminates vapor formation with volatile solvents and is maintenance-free.
The Agilent G1312B SL Binary Pump features a two-channel, dual-piston in-series design which comprises all essential functions of a solvent delivery system. Metering of solvent and delivery to the high-pressure side are performed by two pump assemblies which can generate pressure up to 400 bar. Each channel comprises a pump assembly including pump drive, pump head, active inlet valve, and outlet valve. Both channels are connected in a low-volume mixing chamber which is connected by a capillary coil to a damping unit and a mixer. The design of the 1200 G1312B SL with variable stroke volume allows pulse-free solvent delivery and efficient mixing. An inlet-valve electronically activated and synchronized to piston stroke-eliminates vapor formation with volatile solvents and is maintenance free.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Agilent 1200 HPLC G1312B SL Binary Pump includes:
- > Firmware Version 07.02 (Can be changed upon request)
- > Power Cord
- > Leak and Pressure Diagnostic Report
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Agilent 1200 HPLC G1312B SL Binary Pump Specifications
Hydraulic
system
|
Two dual
piston in series pumps with proprietary servo-controlled variable stroke
drive, floating piston design and active inlet valve
|
Setable flow
range
|
Setpoints
0.001 – 5 ml/min, in 0.001 ml/min increments
|
Flow range
|
0.1 – 5.0
ml/min
|
Flow
precision
|
£ 0.07% RSD,
or £ 0.02 min SD whatever is greater, based on retention time at constant
room temperature
|
Flow accuracy
|
± 1% or 10
µl/min whatever is greater
|
Pressure
|
Operating
range 0 400 bar (0 – 5880 psi) up to 5 ml/min
|
Pressure
pulsation
|
< 2 %
amplitude (typically < 1 %), at 1 ml/min isopropanol, at all pressures
> 1 MPa
|
Compressibility
compensation
|
User-selectable,
based on mobile phase compressibility
|
Recommended
pH range
|
1.0 – 12.5,
solvents with pH < 2.3 should not contain acids which attack stainless
steel
|
Gradient
formation
|
High-pressure
binary mixing, delay volume 180 – 480 µl without mixer, 600 – 900 µl with
mixer, dependent on back pressure
|
Composition
range
|
1 – 99 % or 5
µl/min per channel, whatever is greater
|
Composition
precision
|
£ 0.5%
absolute
|
Composition
accuracy
|
± 0.15% RSD,
at 1 ml/min
|
Control and
data evaluation
|
Agilent
ChemStation for LC
|
Analog output
|
For pressure
monitoring, 2 mV/bar, one output
|
Communications
|
Controller-area
network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down
signals, LAN optional
|
Weight
|
15.5 kg (34
lbs)
|
Dimensions (width
× depth × height)
|
180 x 345 x
435 mm (7 x 13.5 x 17 inches)
|
Electrical
Requirements
|
100 – 240