Agilent 1100 HPLC G1311A Quat Pump
Flow Range 0.2 to 10.0 ml/min
Flow Precision
< 0.3% RSD
Pressure Pulsation
< 2% amplitude
Virtually pulse-free and stable solvent flow with dual floating pistons in series
Variable stroke volume for efficient mixing and consistent solvent delivery
Electronically activated inlet valve synchronized with piston stroke to prevent vapor formation
Settable flow range from 0.001 to 10 ml/min with 0.001 ml/min increments for precise control
Low-pressure quaternary gradient mixing with proprietary high-speed proportioning valve
Wide operating pressure range up to 40 MPa (400 bar) at flow rates up to 5 ml/min
Gradient formation with delay volume of 800 – 1100 µl for optimized performance
Compatible with pH range of 1.0 to 12.5, ideal for various solvents
Compact dimensions of 5.5 x 13.5 x 17 in for efficient bench space utilization
The Agilent 1100 G1311A Quaternary Pump ensures virtually
pulse-free and stable solvent flow, with dual floating pistons in series,
precisely servo-controlled. This design with variable stroke volume allows
pulse-free solvent delivery and efficient mixing. An inlet-valve electronically
activated and synchronized to piston stroke eliminates vapor formation with
volatile solvents and is maintenance free. The Agilent quaternary pump is
equipped with a high-speed proportioning valve.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Agilent 1100 HPLC G1311A Quat Pump includes:
> Firmware Version 07.02 (Can be changed upon request)
> Leak and Pressure Diagnostic Report
> Lan Cable / Power Cord
> Front Cover
> Operator's Manual
> Full 180 Day Parts and Labor Warranty
Agilent 1100 HPLC G1311A Quat Pump Specifications
Setable Flow Range
0.001 – 10 ml/min, in 0.001 ml/min increments
Flow Range
0.2 – 10.0 ml/min
Flow Precision
< 0.3% RSD (usually 0.15%), based on retention time, at 1 ml/min
Pressure
Operating range 0 – 40 MPa (0 – 400 bar, 0 – 5880 psi) up to 5 ml/min
Operating range 0 – 20 MPa (0 – 200 bar, 0 – 2950 psi) up to 10
ml/min
Pressure Pulsation
< 2% amplitude (typically < 1%), at 1 ml/min isopropanol, at
all pressures > 1 MPa (10 bar)
Recommended pH Range
1.0 – 12.5, solvents with pH > 2.3 should not contain acids which
attack stainless steel
Gradient Formation
Low pressure quaternary mixing/gradient capability using proprietary high-speed
proportioning valve
Delay volume 800 – 1100 µl, dependent on back pressure
Composition Range
0 – 95% or 5 – 100%, user selectable
Composition Precision
< 0.2% SD, at 0.2 and 1 ml/min
Dimensions (H x W x D)
5.5 x 13.5 x 17 in
Electrical Requirements
100-120 or 22-240 VAC, 50/60 Hz, 220 VA
Agilent G1311A Operator's Manual
Click Here to View all the Technical Specs for the G1311A Quat Pump