Recertification Protocol
Technical Specification Sheet
 

Agilent 1100 HPLC System with Diode Array Detector & Thermal Auto Sampler


The Agilent 1100 HPLC System guarantees the most efficient mixing and pulse-free solvent delivery for consistently fast and accurate results. The system’s optimized design using a quaternary pump allows variable stroke volumes for stable solvent flow. This electronically activated inlet valve runs synchronous to the pistons to eliminate vapor formation in volatile solvent, making the Agilent 1100 both remarkably easy-to-use and maintenance-free. It also features a high-speed proportioning valve and a pre-configured degasser, which allow convenient low-pressure mixing and a large solvent choice flexibility. The HPLC System’s vacuum degasser ensures reduced operating costs without the use of helium and optimum instrument performance.


The Agilent 1100 Diode Array Detector verifies separation quality with peak purity, multiple signals, and spectral libraries during your analytical quantification, offering an extended wavelength range with the highest sensitivity. The VWD and diode array detectors allow for automated wavelength verification for performance tracking. The Agilent 1100’s temperature management system ensures optimum baseline stability-even under rough ambient conditions. This detector features superb illumination using combined deuterium and tungsten lamps for the highest intensity and lowest detection limit from 190 to 950 nm, 1024 diodes and 1-nm slit for highest spectral resolution. The programmable slit helps you explore the spectral landscape for faster sample characterization, ensuring higher sensitivity or lower baseline noise as necessary on the 1100. Slit settings can be stored and reported from 1 – 16 nm with raw data for GLP traceability.


The unit can be operated with a hand-held controller or computer running Clarity software. The HPLC System features extensive diagnostics with compatible data analysis software for on-site and remote trouble-shooting.


All of our equipment is fully tested to perform at factory specifications before being shipped.


INCLUDED:
The Agilent 1100 HPLC System w/ DAD, ALS Therm includes:
  • > G1311A Quat Pump (Binary Pump Available)
  • > G1329A Autosampler
  • > G1330A ALS Therm
  • > G1315B DAD Detector
  • > G1316A Column Compartment
  • > G1379A Degasser
  • > Windows 10 Professional 64-Bit Data Station with Flatscreen Monitor
  • > Chemstation version 4.3 software installed
  • > Lab Advisor Software B.02.10
  • > Can/Crossover Cables
  • > Agilent G1369 Lan Interface Card
  • > Solvent Tubing
  • > 4 x Bottles, Caps and Solvent Filters
  • > Stainless Steel Capillaries
  • > 1/4" Wrench & Mobile Phase Priming Syringe
  • > Waste Tubing
  • > All Power Cords
  • > Agilent 4 Peak Test w/ Lab Advisor Diagnostics Report
  • > Full 180 Day Parts and Labor Warranty


Below is a List of Diagnostic Tests Performed On This System
G1315B Diode Array Detector Diagnostic Tests:
  • Filter Test: The test moves the filter into the light path and measures the relative absorbance in the filter position.
  • Dark Current Test: The test measures the dark current from the detector optic.
  • Holmium Oxide Test: The test measures the Holmium spectrum from the built-in Holmium filter. The spectrum is evaluated for peaks at different wavelengths.
  • Lamp Intensity Test: The test verifies functionality of the UV lamp at various wavelengths.
  • Wavelength Intensity Test: The test confirms the accuracy of wavelengths measured.
  • Baseline Noise Test: The test measures the amount of baseline noise drift identified by the detector.
G1311A Quaternary Pump Diagnostic Tests:
  • Pressure Test: The pressure test is a quick, built-in test designed to demonstrate the pressure tightness (up to 400 bar) of the system.
  • Leak Test: The leak test is a built-in troubleshooting designed to demonstrate the leak-tightness of the pump.
G1316A Column Compartment Diagnostic Tests:
  • Thermostat Test: Evaluates the cooling and heating performance of the two Peltier elements.
G1329A Autosampler Diagnostic Tests:
  • Injector Steps: Performs each movement of the sampling sequence under manual control. This is useful for confirming autosampler functionality.
System Tests:
  • 4 Peak Test (Isocratic Standard): Standard test published by Agilent to confirm complete functionality of the 1100 HPLC System.

Agilent 1100 HPLC System with DAD Detector Specifications

Type Specification
Detection type 1024-element photodiode array
Light Source Deuterium and tungsten lamps
Wavelength Range 190 - 950 nm
Short term noise (ASTM) Single
and Multi-Wavelength
± 1 × 10-5AU at 254 and 750 nm
Drift 2 × 10-3AU/hr at 254 nm
Linear Absorbance range
> 2 AU (upper limit)
Wavelength accuracy ± 1 nm
Slit width 1, 2, 4 , 8, 16 nm

Diode Width
< 1 nm
Flow Cell

Standard: 13 µl volume, 10 mm cell

path length and 120 bar (1760 psi)

pressure maximum

Control and Data Evaluation Agilent ChemStation for LC
Analog Outputs Recorder/integrator: 100 mV or 1 V,

output range 0.001 – 2 AU, two

outputs

Communications

Controller-area network (CAN), GPIB,

RS-232C, APG Remote: ready, start,

stop and shut-down signals, LAN

optional
Safety and Maintenance Extensive diagnostics, error detection

and display (through control module

and ChemStation), leak detection, safe

leak handling, leak output signal for

shutdown of pumping system. Low

voltages in major maintenance areas.

GLP Features Early maintenance feedback (EMF) for

continuous tracking of instrument

usage in terms of lamp burn time with

user-settable limits and feedback

messages. Electronic records of

maintenance and errors. Verification of

wavelength accuracy with built-in

holmium oxide filter.

Housing All Materials recyclable


Computer Specifications

Model

Dell Optiplex 7050 Desktop PC

Processor

Intel Core i5-6500 Processor 3.6GHz

Ram

16GB

Hard Drive

250GB Internal SSD

Operating System

Windows 10 Professional, 64bit



Agilent 1100 Operators Manual

Click Here to View all the Technical Specs for the Agilent 1100 System
Agilent 1100 HPLC System - Marshall Scientific


Average Rating: Average Rating: 5 of 5 5 of 5 Total Reviews: 1 Write a review »

  7 of 7 people found the following review helpful:
 
5 of 5 Great System October 5, 2020
Reviewer: Rachael Nash from Sacramento, California  
System looks and works great. We configured it yesterday and had no problems. The computer came loaded with the software and included a guide on how to install the system at our location. Steve was very helpful over the phone and pointed us to the right system we required. Thanks again!

Was this review helpful to you?


Browse for more products in the same category as this item:

Chromatography - HPLC
Brand
Brand > Agilent
Chromatography Component
Chromatography Component > Systems