ABI 2720 Thermal Cycler
Thermal Range
4.0°C to 99.9°C
Max Block Ramp Rate 2.7°C/sec
- Designed specifically for the amplification of nucleic acids using the GeneAmp Polymerase Chain Reaction (PCR) process.
- Features an integrated 96-well sample block with an internal Peltier heating/cooling unit for precise temperature control.
- Equipped with cooling vents in the back to facilitate placement in various laboratory setups.
- Easy-to-use graphical representation of methods on the Create/Edit screen for intuitive operation.
- Users can create and store their most commonly used methods for quick access and efficiency.
- Provides the ability to lock methods, safeguarding them from unauthorized changes. Only users with the correct PIN can overwrite or delete locked methods.
The Applied Biosystems 2720 Thermal Cycler is an automated instrument, specifically designed for the amplification of nucleic acids using the GeneAmp Polymerase Chain Reaction (PCR) process. The ABI 2720 Thermal Cycler provides the same reliability and performance that customers around the world have come to expect from Applied Biosystems thermal cyclers.
The instrument has an integrated 96-well sample block, which houses an internal Peltier heating/cooling unit. The sample block is made of aluminum to provide optimal thermal transfer rates. The ABI 2720 Thermal Cycler's exceptionally small footprint enables it to fit almost anywhere. With cooling vents in the back as well, this enables it to fit in tight spaces to save valuable bench space.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The ABI 2720 Thermal Cycler includes:
- > 96 Well Block
- > Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
ABI 2720 Thermal Cycler Specifications
Thermal Range
|
4.0°C to 99.9°C
|
Accuracy
|
±0.5°C (Range: 35° – 99.9°C)
|
Thermal Uniformity
|
±0.5°C (30 sec. after reaching 95°C)
|
Ramping Rate
|
Max Block Ramp Rate: 2.7°C/sec
Max Sample Ramp Rate: 2.6°C/sec
|
Sample Capacity
|
96 x 0.2 ml Tubes or 96-Well Plate
|
Memory
|
Stores 100 methods (Pre/Post-PCR Holds and Cycles)
|
Gradient Function
|
No
|
Dimensions (W x D x H)
|
8.3 x 14.2 x 8.7 in (21 x 36 x 22 cm)
|
Weight
|
13.5 lbs (6.1 kg)
|
Electrical Requirements
|
100-240V, 50/60Hz
|
ABI 2720 Thermal Cycler Manual
Click Here to View the Technical Specs for the ABI 2720 Thermal Cycler
- Validate that the power supply and PCB consistently deliver accurate, stable voltages across all circuits and connections, confirming operational reliability
- Thoroughly inspect the display for any dim, flickering, or dead pixels, replacing compromised segments to ensure clear visibility and precise data display
- Verify complete functionality and responsiveness of all buttons and knobs, confirming accurate and reliable input across all controls
- Ensure the machine completes all internal diagnostic checks with zero errors, indicating system readiness and functional integrity
- Confirm the machine accurately identifies attached blocks (96-well, dual 48, etc.), with seamless recognition across different configurations
- Conduct a detailed inspection of internal wiring and all electrical connections, checking for any signs of breaks, disconnections, or tears; replace or repair any underperforming components
- Examine the connector between the block and base for bent or broken pins, replacing as needed to maintain secure, reliable connections
- Calibrate well temperatures with a NIST-traceable thermometer, verifying accuracy and uniformity across all wells to meet specification standards
- Perform a comprehensive cosmetic inspection, ensuring all exterior surfaces and components are free of damage, wear, or defects, maintaining aesthetic and professional quality
Technician Inspection and Recertification Checklist: Thermal Cycler
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, accurate voltage outputs from power supply and PCB across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed for clarity
|
Completed
|
|
Verify functionality and responsiveness of all buttons and knobs
|
Completed
|
|
Confirm machine completes internal diagnostics with zero errors
|
Completed
|
|
Ensure accurate block identification across configurations (96-well, dual 48, etc.)
|
Completed
|
|
Inspect internal wiring and connections for integrity; repair or replace as needed
|
Completed
|
|
Check connector between block and base for bent or broken pins; replace as necessary
|
Completed
|
|
Calibrate well temperatures with a NIST-traceable thermometer for accuracy
|
Completed
|
|
Conduct cosmetic inspection to ensure quality and professionalism of all surfaces
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|