YSI 2300 STAT Plus Glucose Lactate Analyzer
The YSI 2300 STAT Plus is the only dedicated dual-channel glucose
and lactate analyzer on the market, providing exceptional speed, performance,
and ease of use. This analyzer measures glucose in blood, plasma, and serum and
lactate in blood, plasma, and cerebrospinal fluid. The YSI 2300 STAT Plus
lowers cost per test with no need to renew enzymes, saves time through whole
blood capability with no need to spin blood first, and allows batch runs up to
24 samples. This system also features interference rejection and a broad measurement
range, which eliminates the need for sample dilution.
Standard features include: microprocessor control; menu-driven setup with battery backed memory; alpha-numeric liquid crystal display; built-in data print; RS-232 serial port installed; turntable interface port installed; complete diagnostic software; automatic calibration; programmable calibration protocol; selectable performance modes; selectable concentration units of mmol/L, mg/dL, mg/L; direct measurement of glucose in whole blood, plasma, and serum; direct measurement of lactate in whole blood, plasma, and CSF; sample aspiration of only 25 microliters; sample tube holder for "hands off" sampling; manual sampling station for aspirating small volumes; and conversion of whole blood to plasma glucose values (user enters hematocrit value at keypad).
The YSI 2300 STAT Plus is ideal for use in the following:
STAT labs; diabetes, human performance, and exercise and athlete evaluation;
diabetes and endocrinology, general metabolic, and veterinary medicine research;
handheld monitor development; cardiac and pulmonary rehabilitation; neonatology
and pediatrics; and pharmaceutical quality assurance.
Three in stock at various ages, please inquire.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The YSI 2300 STAT Plus Glucose Lactate Analyzer includes:
- > Calibration Report
- > YSI 2747 Glucose Cal Solution
- > YSI 2357 Buffer Concentrate
- > YSI 2751 Printer Paper
- > Operators Manual
- > Full 180 Day Parts and Labor Warranty
YSI 2300 STAT Plus Glucose Lactate Analyzer Specifications
Sample Size
|
25 uL (aspirated volume)
|
Response Time
|
Normal Mode: Display and
print results in 65 seconds. Sample to sample interval is 100 Seconds
Screen Mode: Display and
print results in 45 seconds. Sample to sample interval is 70 seconds
|
Measurement Range
|
Normal Mode:
Glucose: to 900 mg/dL (9000 mg/L, 50.0
mmol/L)
Lactate: to 30.0 mmol/L (267
mg/dL, 2670 mg/L)
Screen Mode:
Glucose: to 500 mg/dL (5000 mg/L, 27.8 mmol/L)
Lactate: to 15.0 mmol/L (134 mg/dL, 1335 mg/L)
|
Precision
|
Glucose: 2% of reading or 2.5 mg/dL
Lactate: 2% of reading or 0.1 mmol/L
|
Resolution
|
Glucose: 1 mg/dL (1 mg/L, 0.1 mmol/L)
Lactate: 0.1 mmol/L (1 mg/L, 1 mg/dL)
|
Membrane Working Life (typical)
|
Glucose: 21 days
Lactate: 14 days
|
Operating Temperature
|
15 to 35
Celcius
|
Humidity
|
10-90% non-condensing
|
Power
|
110-120 VAC or 220-240 VAC, 50-60 Hz
|
Dimensions/ Weight
|
25.4 x 35.6 x 35.6 cm/ 11.4 kg
|
Interfacing
|
Serial port for RS 232C
Optional bar code reader
|
Regulatory Compliance
|
CE, FDA Registered Class II, CSA
|
YSI 2300 Operator's Manual